This device, based on electrochemical impedance spectroscopy (EIS), enables continuous monitoring of TS sensor impedance by applying voltages at two specific frequencies.

Logger Lineup

Corrosion Monitor

  • Measure up to 2 sensors.
  • Battery or AC adapter powered.

Custom specifications available to suit your environment.
For any requests, please contact us

Specifications

Atmospheric corrosion systems under electrolyte immersion and a water film are represented by the equivalent circuit:
Z: Impedance
Rp: Polarization resistance (∝ corrosion resistance)
Cdl: Electric double layer capacitance
Rs: Solution resistance

Since it can be approximated that:

  • Z ≒ Rs when ω→∞
  • Z ≒ Rs + Rp when ω→0

The solution resistance Rs can be measured from the high frequency side, and the polarization resistance Rp can be calculated by subtracting the value on the high frequency side from the value on the low frequency side.

The inverse of the polarization resistance is correlated with the corrosion rate. Therefore, it is possible to grasp the change in the corrosion rate over time.

Number of Measurement Channels

SICM-714B: max.4
SICM-718B: max.8


Data Acquisition and Format

10 mins interval (Adjustment Possible)


Applied voltage

Sine wave, Select 20mVp-p, or 100mVp-p


Impedance measurement range

High frequency ~10 5 Ω, Low frequency ~10 9 Ω


How to measure

AC impedance measurement with DFI


Noise reduction

Noise cancellation by integrating 1 to 8 cycles,
Noise removal with shield/guard function,
Cable stray capacitance cancellation


Data Acquisition and Format

10 mins interval (Adjustment Possible)


Recorded data

Time, Solution resistance“Rs ”, Corrosion resistance “Rc ”,
Electric double layer capacitance “Cdl ”, Phase difference


Data Acquisition and Format

Text format (readable by spreadsheet software, etc.)


Data storage

Industrial SD card


Power

100V AC / 12V Battery


Weight

SICM-714B: Approx.1.7kg
SICM-718B: Approx.1.9kg


Size

260mm×180mm×100mm (Excluding protrusions)